The DOP2000 uses pulsed ultrasonic
echo technique to measure instantaneously velocity profiles,
Doppler energy profiles, echo profiles, spectral density
(FFT), histogram and the flow rate in flowing liquids.
It computes and displays in real time these data profiles
based on the analysis of a user¡¯s specified number of
gates placed along the ultrasonic beam.
The digital ultrasonic synthesizer included can generate
any emitting frequencies between 0.45 MHz and 10.5 MHz.
Associated to this performance, the DOP2000 includes a
variable spatial resolution filter that allows to adapt
the size of the sampling volume to the application and
therefore improves the signal to noise ratio of the measurements.
All the ultrasonic parameters (frequency, Prf, Tgc..)
and the processing conditions (number of gates, filters...)
are set by the user. The smart trigger capability of the
instrument allows to synchronize the acquisition to any
periodic or non periodic event. This high flexibility
applies to any of the 10 channels multiplexer (available
in option) and up to 32¡¯000 profiles could be recorded
in binary or ASCII format. If desired, the DOP2000 could
record simultaneously two types of data profiles, such
as the velocity profiles and the Doppler energy profiles.
A separate emitter output connector enables to use two
different transducers for emission and reception.
Operating in a Windows¢ç9x environment, the velocimeter
can be adapted to your applications with only few mouse
clicks. The measured profiles are displayed on a bright
800x600 TFT color screen and are recorded in its internal
memory or send through the build-in Ethernet connection
to any storage device within few milliseconds. |
| |
| |
[Á¦Ç°»ç¾ç]
Emission
| Emitting frequencies |
from 0.45 MHz to 10.5 MHz, step of
1 KHz |
| Emitting power |
3 levels |
| Burst length |
2, 4 or 8 cycles |
| PRF |
between 64 ¥ìs and 10¡¯500 ¥ìs, step
of 1 ¥ìs |
Reception
| Number of gates |
between 4 and 1000, step of 1 gate |
| Amplification (TGC) |
slope mode, custom mode, variable
level between -40 and +40 dB, step of 1 dB |
| Sensitivity |
5 levels, maximum -100 dBm |
Resolution
| Lateral resolution |
defined by the transducer |
| Longitudinal resolution |
minimum 0.8 ¥ìs, 0.6mm (-3 dB) |
| Spatial filter |
from 50 KHz to 300 KHz, step of 50
KHz |
| Time between gates |
variable, between 250 ns and 20 ¥ìs,
step of 250 ns |
| Velocity resolution |
1 LSB, result given in a byte format |
Ultrasonic processor
| Emissions / profile |
between 8 and 512, any value |
| Acquisition time/profile |
variable, minimum value 2 ms |
| Real time filters |
moving average, median |
| Velocity scale |
maximum: ¡¾11 m/s down to ¡¾1mm/s,
sliding origin |
Computation
| Curve displayed |
velocity profile
Doppler energy
echo modulus
velocity versus time
flow versus time
histogram
power spectrum of a gate |
| Statistics |
mean value, standard
deviation, histogram |
| Utility |
freeze/run mode
2 cursors |
Trigger
| Input |
external signal (TTL) or keyboard
action |
| Condition |
high, low level, internal pull-up
4 KW |
| Delay |
between 1 and 10¡¯000 ms, step of
1ms |
| Acquisition procedure |
selectable number of blocks of profiles |
Memory
| Size |
Up to 32¡¯000 profiles,
step of 1 profile |
| Configuration |
10 saved configurations |
Environment
| Operating system* |
Windows 95 or 98 |
| Processor* |
Pentium MMX, 233 MHz |
| RAM* |
128 Mbytes, optional 256 MBytes |
| Storage devices* |
hard disk 20 GBytes
1,44 MBytes Floppy, CD-ROM 44x or CD-RW 2/4/32 in
option |
| Screen |
12.4¡¯¡¯ TFT Color display (800x600)
VGA |
| Communication |
2 serial ports
1 parallel port
1 ethernet 10 base T, (RJ45)
1 SVGA port (simultaneous)
1 keyboard (miniDIN) |
| US interface signals |
echo (max 0.7 Voltp), BNC
synchro pulse (TTL), BNC
trigger input, TTL level, BNC
probe in/out, BNC
probe out, BNC |
| Power supply |
220-110 VAC selectable, 50-60 Hz |
| Humidity |
=< 80% |
| Temperature |
5-35 ¡ÆC |
| Size |
340x265x305 cm |
| Weight |
13 kg |
Options
| Multiplexer |
10 probes, internal
or external |
| Sound speed
unit |
measure the sound velocity
within 2% |
* may be adapted to the market
All values computed with a sound velocity of 1500 m/s
(water), in the direction of the ultrasonic beam
|
| |
|
| |
|
| |
|
|